METROLOGY - PRESSURE CAPABILITIES

PARAMETER

UNITS

RANGE

ACCURACY

TECHNIQUE


PRESSURE HYDRAULIC

Pounds per Square Inch

20 to 10,000 psi ± 0.005% IV Standard DH Deadweight Testers

(psi)

10,001 to 30,000 psi ± 0.01% IV  
  30,001 to 50,000 psi ± 0.02% IV  


PRESSURE PNEUMATIC

Pounds per Square Inch Absolute (psia)

Vacuum to 87 psia ± 0.005% FS + .005% IV DH Digital Piston Gage

Pounds per Square Inch

1.5 to 8,000 psi ± 0.005% IV Standard DH Deadweight Testers

(psi)

8,001 to 16,000 psi ± 0.01% IV  

Inches of Water

0 to 2 In. H2O ± 0.001 In. H2O Dwyer Electronic Point Gage

(In. H2O)

0 to 24 In. H2O ± 0.005 In. H2O Dwyer Hook Gage
  1 to 60 In. H2O
(0.036 to 2.2 psi)
± 0.005% FS + .005% IV DH Digital Piston Gage

pascals (pa)

2 Kpa to 350 Kpa
gauge mode
± 0.0035% of reading or 0.5 pa, whichever is greater DH Digital Piston Gage
  5 Kpa to 350 Kpa
absolute mode
   
NOTES:
1
. Calibration of deadweight tester by the cross-float technique to determine the effective area for a piston-cylinder combination.
2. Calibration of deadweight tester to verify the existing mass to pressure conversion coefficient.
3. Listed accuracies are per manufacturer's specifications.