METROLOGY - PRESSURE CAPABILITIES |
||||
|
PARAMETER |
UNITS |
RANGE |
ACCURACY |
TECHNIQUE |
|
|
Pounds per Square Inch |
20 to 10,000 psi | ± 0.005% IV | Standard DH Deadweight Testers |
|
(psi) |
10,001 to 30,000 psi | ± 0.01% IV | ||
| 30,001 to 50,000 psi | ± 0.02% IV | |||
|
|
Pounds per Square Inch Absolute (psia) |
Vacuum to 87 psia | ± 0.005% FS + .005% IV | DH Digital Piston Gage |
|
Pounds per Square Inch |
1.5 to 8,000 psi | ± 0.005% IV | Standard DH Deadweight Testers | |
|
(psi) |
8,001 to 16,000 psi | ± 0.01% IV | ||
|
Inches of Water |
0 to 2 In. H2O | ± 0.001 In. H2O | Dwyer Electronic Point Gage | |
|
(In. H2O) |
0 to 24 In. H2O | ± 0.005 In. H2O | Dwyer Hook Gage | |
| 1 to 60 In. H2O (0.036 to 2.2 psi) |
± 0.005% FS + .005% IV | DH Digital Piston Gage | ||
|
pascals (pa) |
2 Kpa to 350 Kpa gauge mode |
± 0.0035% of reading or 0.5 pa, whichever is greater | DH Digital Piston Gage | |
| 5 Kpa to 350 Kpa absolute mode |
||||
| NOTES: 1. Calibration of deadweight tester by the cross-float technique to determine the effective area for a piston-cylinder combination. 2. Calibration of deadweight tester to verify the existing mass to pressure conversion coefficient. 3. Listed accuracies are per manufacturer's specifications. |
||||