<< Back to Electrical Lab

METROLOGY - DC/LOW FREQUENCY CAPABILITIES

PARAMETER

UNITS

RANGE

ACCURACY/
UNCERTAINTY

TECHNIQUE

REMARKS

VOLTAGE DC

Volt (V)

Reference Zeners

Intercomparison to reference group.  
10.00 V 0.4 ppm
1.018 V 0.5 ppm
1.000 V 0.5 ppm
    100.0 mV 2.6 ppm Comparison utilizing transfer techniques to various voltage levels. Meters and Multifunction Calibrators
1.0 V 1.1 ppm
10.0 V 1.1 ppm
100.0 V 1.1 ppm
1000.0 V 1.2 ppm
    > 1.1 kV to 50 kV ± 0.013% IV Measurements made via a precision high voltage resistive divider.  

VOLTAGE AC

Volt (V)

10 Hz to 1 MHz:

(IV)

Thermal Transfers Volt-Hertz Product:
1 x 10
8 @ 100 kHz
2.2 x 10
7 @ 1 MHz
.002 V to 0.02 V ± 0.0170% to ± 0.5382%
> 0.02 mV to 0.2 V ± 41 ppm to ± 0.1045%
> 0.2 V to 0.6 V ± 30 ppm to ± 0.0311%
> 0.6 V to 2 V ± 15 ppm to ± 0.0208%
> 2 V to 6 V ± 13 ppm to ± 0.0195%
> 6 V to 20 V ± 18 ppm to ± 0.0193%
> 20 V to 60 V ± 25 ppm to ± 0.0193%
   

10 Hz to 300 kHz:

     
> 60 V to 200 V ± 22 ppm to ± 0.0195%
   

10 Hz to 100 kHz:

     
> 200 V to 1000 V ± 29 ppm to ± 0.0194%
   

@ 60 Hz:

  Measurements made using a precision high voltage divider and a digital voltmeter.  
> 1000 V to 50 kV ± 0.25%
© Copyright 2000 Edison ESI